Polychromatic Beer-Lambert Analysis Allura Red × 510/10 nm Filter — Measured Nanodrop Spectra
Repeats the analysis from filter_update_analysis.py ("Allura Red × 510/10 Filter" section)
using directly measured Nanodrop UV-Vis spectra recorded 2026-08-31, replacing the
digitised + ε₅₀₈-scaled spectral shape.
2026-08-31 · Allura Red AC (FD&C Red 40) · 10 mM HEPES blank · Nanodrop L = 1.0 mm
1 · Measured Spectra Overview
Five Allura Red concentrations (0.375, 1.0, 1.5, 2.5, and 5 mM) were measured in the Nanodrop
against a 10 mM HEPES blank at a 1.0 mm path length. The measured visible absorption peak is
493–496 nm, slightly blue-shifted from the ~506 nm value in the digitised spectrum
used previously. The 510/10 filter passband (480.8–536.9 nm) overlaps this peak well.
Figure 1.
All five measured Allura Red Nanodrop spectra (L = 1.0 mm, 10 mM HEPES blank) overlaid with the
normalised 510/10 nm filter transmittance (right axis, purple). Concentrations above ~0.375 mM
show saturation artefacts at the visible absorption peak.
2 · Reference Spectrum Selection
The polychromatic calculation requires A(λ) to be in the Beer-Lambert linear
regime. At L = 1.0 mm the expected peak absorbance is approximately:
Concentration
Expected A₅₀₈ (L=1.0 mm)
Status
0.375 mM
~0.97
✓ linear
1.0 mM
~2.59
⚠ moderate saturation
1.5 mM
~3.88
✗ saturated
2.5 mM
~6.47
✗ severely saturated
5.0 mM
~12.9
✗ off-scale
The 0.375 mM spectrum is used as the reference.
It has the lowest saturation risk in the 480–540 nm passband while still providing good signal.
All higher concentrations are used only for the Beer-Lambert linearity check below.
Figure 2.
Allura Red reference absorbance spectrum at 0.375 mM (L = 1.0 mm, left axis, green) overlaid
with the 510/10 filter transmittance (right axis, purple). The filter passband (480–537 nm)
sits across the visible absorption peak, maximising signal.
3 · Beer-Lambert Linearity Check
The filter-weighted polychromatic absorbance A_poly is computed for each measured
concentration and compared against the ideal linear prediction anchored at 0.375 mM:
Figure 3.
Filter-weighted absorbance A_poly for each measured concentration at L = 1.0 mm,
versus the ideal Beer-Lambert line anchored to 0.375 mM (dashed). Strong negative deviations at
1.0 mM and above confirm Nanodrop saturation; only the 0.375 mM point is usable as a spectral
reference.
Concentration (mM)
A_poly (measured)
Measured ratio / 0.375 mM
Expected ratio
Deviation (%)
0.375
0.6947
1.0000
1.0000
0.00 %
1.0
1.6614
2.3914
2.6667
−10.32 %
1.5
2.1795
3.1371
4.0000
−21.57 %
2.5
2.3458
3.3764
6.6667
−49.35 %
5.0
1.8024
2.5944
13.3333
−80.54 %
4 · Polychromatic Beer-Lambert Analysis
Using the 0.375 mM reference spectrum, the absorbance at any working condition
(C, L) is scaled as:
4.1 · Polychromatic transmittance at the reference condition
Reference condition: 0.375 mM · L = 1.0 mm
Filter-weighted mean A = 0.7022
T_poly (integrated) = 0.20196 → A_meas = 0.69475
T_naive (10−Ā) = 0.19853 → A_mean = 0.70217
Polychromatic underestimate: 1.057 % — negligible for this dye–filter pair.
Figure 4.
Monochromatic transmittance T(λ) across the 510/10 passband (blue fill) versus the
polychromatic average T_poly (red dashed) and the naïve single-wavelength approximation
T_naive (green dotted). The near-flat transmittance profile across the passband is why the
polychromatic error is so small (<1.1 %).
4.2 · Absorbance vs. concentration fraction and deviation ratio
Scaling to a working concentration of 1.0 mM in the channels and a capillary calibration
path length of 80 µm, the polychromatic absorbance is computed at four concentration fractions
of the 1.0 mM stock:
Figure 5.Left: Polychromatic A_meas (tomato) vs. naïve Beer-Lambert A_naive (green) at the
80 µm capillary for four concentration fractions of the 1.0 mM stock.
Right: Deviation ratio A_poly / A_naive — values within 0.2 % of unity confirm that
the polychromatic correction is negligible for Allura Red × 510/10.
4.3 · Effective molar absorptivity
Figure 6.
Filter-weighted effective molar absorptivity ε_eff (tomato) versus true monochromatic ε (green
dashed) for the four concentration fractions at L = 80 µm. The near-constant ε_eff = 1.871
L mol⁻¹ µm⁻¹ confirms Beer-Lambert linearity in the measurement range.
5 · Polychromatic A_meas Table
Computed A_meas for all eight channel heights (30–200 µm) at four concentration fractions of the
1.0 mM working stock. Reference calibration: 0.375 mM Nanodrop spectrum at L = 1.0 mm, scaled
to (C, L) via Beer-Lambert.
Height (µm)
f = 1/8 (0.125 mM)
f = 1/4 (0.250 mM)
f = 1/2 (0.500 mM)
f = 1/1 (1.000 mM)
30
0.0070
0.0140
0.0281
0.0561
40
0.0094
0.0187
0.0374
0.0748
50
0.0117
0.0234
0.0468
0.0935
70
0.0164
0.0328
0.0655
0.1309
90
0.0211
0.0421
0.0842
0.1682
120
0.0281
0.0561
0.1122
0.2241
150
0.0351
0.0702
0.1402
0.2799
200
0.0468
0.0935
0.1868
0.3726
All values in absorbance units (AU). Working concentration in channels: 1.0 mM.
Capillary calibration path length: 80 µm.
6 · Height Estimation Accuracy
Heights are estimated from measured A_meas using the capillary calibration at 80 µm:
h_est = 80 µm × A_meas(h) / A_meas(80 µm).
The error (h_est − h_true) / h_true quantifies the residual polychromatic bias.
Figure 7.Left: Estimated height vs. true channel height for all four concentration fractions
(plasma colormap). All curves lie on the ideal line (dashed), confirming negligible bias.
Right: Height estimation error — all values are within ±0.2 % for all heights and
concentrations, confirming that the polychromatic correction is not needed for this
dye–filter combination.
Key conclusion:
The polychromatic Beer-Lambert error for Allura Red × 510/10 is < 0.2 %
across all channel heights (30–200 µm) and concentration fractions (1/8–1/1 of 1.0 mM).
The height estimation is essentially linear, and no polychromatic correction is needed.
This confirms the result from the digitised-spectrum analysis — the direct Nanodrop measurement
produces consistent conclusions.
7 · Notes on Nanodrop Measurement Quality
Peak wavelength: The measured visible absorption peak is at 493–496 nm,
slightly blue-shifted from the ~506 nm value in the digitised reference spectrum.
This difference may reflect solvent composition, pH, or digitisation uncertainty in the
literature source.
Saturation at L = 1.0 mm: Only the 0.375 mM spectrum is in the linear
Beer-Lambert regime. For future Nanodrop measurements, use ≤ 0.4 mM or the shorter
0.1 mm Nanodrop path if available.
NaN values: The 2.5 mM and 5 mM spectra have NaN readings (5 % and 20 %
respectively) in the 480–540 nm passband due to detector saturation.
These spectra are not used for the polychromatic calculation.
Effective ε: ε_eff = 1.871 L mol⁻¹ µm⁻¹ from the measured spectrum versus
2.59 × 10⁴ L mol⁻¹ cm⁻¹ = 2.59 L mol⁻¹ µm⁻¹ from the literature value used previously.
The 28 % difference likely reflects the digitisation error in the literature spectrum shape
and/or concentration uncertainty in the 0.375 mM Nanodrop sample.
The capillary calibration strategy eliminates this uncertainty from the height inference.